Meriam M4 Series Precision Calibrator Data Logger Manual Instrukcja Użytkownika Strona 32

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 40
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 31
ZM4 Instruction Manual 9R87-B June 2012
32
Specifications - Over and Under Range
Soft Under & Over Range:
Indicates pressure is outside of the accuracy specification of the sensor. RANGE indicator will
appear.
Hard Under & Over Range:
Indicates pressure is significantly outside of the measurement range of the sensor. RANGE
indicator will appear. Display will read “-----“. Red backlight will illuminate.
Certified Range:
Sensor is operating within the range of its accuracy specification.
NOTE:
The over/under range threshold is set in the sensor. The value, as shown above, is typically set
to +/-20% of Full Scale.
Przeglądanie stron 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 38 39 40

Komentarze do niniejszej Instrukcji

Brak uwag